Troubleshooting and Maintenance Guide
10
General Maintenance
To Reinstall the EI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
Slide the ion source into the source radiator (
).
2
Install and hand tighten the source thumbscrews. Do not
overtighten the thumbscrews.
3
Connect the ion source wires as shown in
4
Close the analyzer chamber.
5
Pump down the instrument. See the
5975T LTM GC/MSD
Operation Manual.