Trouble shooting
Operator’s Manual
7.1.2
C-codes.
C-codes are offered to qualified Service Technicians
only.
Factory calibration not completed.
•
Call Technical Service.
E51
Calibration failed
•
Switch unit off, wait for 20 seconds.
•
Switch unit on.
•
Retry calibration, or:
•
Call Technical Service.
E52
Calibration weight on opposite side respect the factory
calibration.
•
Attach the User Calibration Weight correctly on the
left side of the flange. Redo Calibration.
•
Call Technical Service.
•
Call Accu Service.
E51
Calibration failed
•
Switch unit off, wait for 20 seconds.
•
Switch unit on.
•
Retry calibration, or:
•
Call Accu Service.
E52
Calibration weight on opposite side to factory
calibration.
•
Attach the User Calibration Weight correctly on the
left side of the flange. Redo Calibration.
•
Call Accu Service.