![background image](http://html1.mh-extra.com/html/abbott/architect-rhs/architect-rhs_training-manual_450972085.webp)
MODULE 3: ROBOTICS
ARCHITECT
®
RSH System Service Training Guide
3-11
201532-102
FOR INTERNAL USE ONLY
Diagnostic Information:
•
Carrier Transport must be removed and installed at the extreme left end behind Bar Code
Reader to avoid damaging Section Align Sensor
•
M&D 1119 Transport Calibration Procedure aligns the Carrier Transport position with
positioner pockets, tray and Carrier positions
•
Carrier Transport Belt must contain a minimum of 10 evenly spaced cords
•
Bearing felt wiper pads are replaceable and should be super saturated with SuperLube oil
•
Proper Belt positioning is important for proper operation
•
J702 and J701 communication cables are identical cables and can be mispositioned during
installation to create errors
•
Carrier Transport Flex Board de5VDC to the Rail Guide Sensor Board
•
Motors receive 36VDC from the RSH Card Cage
•
Rail Guide Sensor Board is replaceable