MODULE 5: MISCELLANEOUS
5-10
ARCHITECT
®
RSH Service Training Guide
FOR INTERNAL USE ONLY
201532-102
Processing Module Calibrations
Calibrate Processing Module Pipettors.
Procedure
M or D
FUNCTION and USE
Calibrates
1111 Sample
Pipettor Calibration
M
For
i
2000
SR
®
System ONLY
• Probe Straightness is determined.
• Positions are set for Sample Carriers, or
LAS Sample Carousel Wash station, RV1
and RV24.
• Sample Pipettor to Wash Cup
• Process Path to Position 1
• Process Path to Position 24
• Aspiration Location at RSH
1117 STAT Pipettor
Calibration
M
For
i
2000SR
System ONLY
• STAT positions are set for Sample
Carriers, Wash Station, and RV 47.
• STAT Pipettor Wash Cup
• Process Path Position 47
• STAT aspiration location at RSH
1120 Sample
Pipettor Calibration
M
For
c
8000
®
System ONLY
• Sample Pipettor Calibration.
• automatically determines the position of
calibration targets in relation to the home
position of the pipettor.
• Sample Handler Carrier Position
• Reaction Carousel
• Sample Carousel
• Wash Cup
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...