![background image](http://html1.mh-extra.com/html/abbott/architect-rhs/architect-rhs_training-manual_450972123.webp)
MODULE 4: POWER AND CIRCUIT BOARDS
ARCHITECT
®
RSH Service Training Guide
4-21
201532-102
FOR INTERNAL USE ONLY
M&D PROCEDURES
M & D procedures are grouped in two separate screen sections under the SYSTEM Menu, Diagnostics and
Maintenance. The procedures in each section are grouped by module type, and then by category. The
A
RCHITECT RSH Service and Support Manual
provides guidance regarding module, category, user
logon, time required, along with a brief description of each procedure.
The DIAGNOSTIC screen displays the available diagnostic procedures. Depending on user logon,
available Diagnostic Procedures vary. Each category is represented by a tab on the Diagnostics screen.
Once a procedure is initiated, step-by-step instructions walk the user through completion.
Maintenance procedures are either scheduled or unscheduled. Each maintenance category is represented
by a tab on the screen. Once a procedure is initiated, step-by-step instructions walk the user through
completion. The user can also access windows to view version and detail information for each procedure
and print the procedure report.
NOTE: For additional information regarding M&D procedures refer to the
ARCHITECT RSH System Service and Support Manual and the ARCHITECT
System Operations Manual.
Below is a listing of M&D Procedures related to Power and their function:
Procedure
M or D
FUNCTION and USE
When to Perform
6008 Controller
Configuration
D
• Used to configure the PM and RSH controller
board after replacement. Bootcode is
downloaded to the RSH or PM CPU. This
procedure configures the module so new
bootcode will be downloaded into memory
during the next power cycle.
A power cycle must be performed for changes to
take effect.
• Install
• Removal and Replacement
5501 RSH Motor Test
D
• For RSH only
Used to home the following RSH motors:
Carrier Transport X, Carrier Transport theta,
Carrier Transport Z, and Carrier Positioner.
• Troubleshooting
5506 RSH Sensor Test
D
• For RSH only
Use this procedure to test the Carrier Transport
and Carrier Positioner Sensors.
• Troubleshooting
5507 RSH Indicator
Lights Test
D
• For RSH only
Use this procedure to test the RSH indicator
lights.
• Troubleshooting
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...