MODULE 5: MISCELLANEOUS
ARCHITECT
®
RSH System Service Training Guide
5-11
201532-102
FOR INTERNAL USE ONLY
Diagnostic Tips:
•
Pipettor calibration failures can be caused by:
-
LLS Antenna problems: poor connections at the antenna or card cage, LLS antenna
failure
-
LLS Board problems: poor connection of the board in the card cage, board failure
-
STAT
i
Sample only, sample handler problems: carrier is not being moved to the
aspiration point correctly.
-
Calibration targets not cleaned
-
Probe failure
-
RSH attached incorrectly to PM
-
Left to right position is OFF
-
Reference line on
i
2000
SR
®
System or leveling procedure incorrectly
performed
•
Transport calibration failures can be caused by:
-
Poor cable connections at the transport or SIB board
-
Carrier Transport failure
-
Slot damaged or obstructed
-
Obstruction in Carrier Transport movement
•
RSH Bar Code Reader calibration failures can be caused by:
-
RSH Bar Code Reader failure
-
Incorrect positioning of calibration tools
-
Poor cable connections at the Bar Code Reader, SIB, or Card Cage
-
Carrier Transport movement problem
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...