MODULE 4: POWER AND CIRCUIT BOARDS
ARCHITECT
®
RSH Service Training Guide
4-27
201532-102
FOR INTERNAL USE ONLY
KEY POINTS
In this module, you learned about power, power distribution and boards. You were familiarized with:
•
Power Distribution
•
Power distribution varies depending on instrument configuration
-
c
8000
®
System SMD/SH Power Supply provides power to RSH Card Cage
-
i
2000
SR
®
and
ci
8200
®
Systems use the
i
2000
SR
Power Supply and Card Cage for power
distribution
•
If the RSH Status remains OFFLINE and there is no power, this may be the result of:
-
200VAC failure or malfunction from Main Power Supply on the
c
8000
System
-
Loose power cables at RSH or card cage
-
Failed fuses at RSH boards
•
Three motors receive 36VDC:
-
X Motor
-
Transport Carrier Assembly (Z and theta motors)
-
Carrier Positioner Assembly
•
Card Cage
•
Card Cage configuration varies depending on instrument configuration
-
i
2000
SR
System and
ci
8200 Systems use the
i
2000
SR
System Card Cage
-
JU1 and JU2 jumpers installed on the Card Cage backplane indicate that the
system is configured for the RSH
-
c
8000 System the RSH Card Cage is a separate Card Cage from the main
c
8000 System.
It is attached to the right side of the Processing Module.
-
JU1 and JU2 jumpers installed on the Card Cage backplane indicate that the
system is configured for the RSH
•
Circuit boards
-
1.8 and 3.3V is developed and used on the SIB
-
Sample Handler Controller functions as CPU for Sample Handler
-
Indexer Board controls and tracks motor movement
-
Motor Driver Board provides motor drive
-
i
2000
SR
and
ci
8200
Systems contain multiple Indexer, Motor Driver and Controller
(CPU) boards that are identical and therefore can be swapped for troubleshooting
purposes
-
UDB contains switches that must be set during board installation
-
Controller Board contains two reset switches for an alternate method to reboot and
initialize the RSH
-
When installing new Controller Boards, M&D 6008 Controller Configuration is required
for downloading new boot code data to board
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...