![background image](http://html1.mh-extra.com/html/abbott/architect-rhs/architect-rhs_training-manual_450972168.webp)
APPENDIX
A-14
ARCHITECT
®
RSH Service Training Guide
FOR INTERNAL USE ONLY
201532-102
1.
In the table below, fill in the Voltage supplied by the RSH Card Cage to the indicated device/
component:
2.
What voltages are developed and used on the SIB Board?
3.3VDC and 1.8VDC
3.
If +5VDC power was not present at the SIB which components might be affected?
RSH Bar Code Reader
LUI Keypad
PLSB and UDB
Carrier Positioner Board
Carrier Transport Sensors
4.
On a
ci
8200
®
System which Card Cage controls the RSH?
i2000
SR
®
System Card Cage
5.
What boards are contained in the RSH Card Cage on a
c
8000
®
Standalone System?
Module Controller or CPU
Indexer
Motor Driver
Component
Voltage
SIB
5
Carrier Transport X Motor
36
Indexer
5,+ 12