MODULE 2: BASIC OPERATION
2-18
ARCHITECT
®
RSH Service Training Guide
FOR INTERNAL USE ONLY
201532-102
Routine Bay (Loading Platform)
All non-priority load/unload points on the RSH are referred to as Routine Bays. A tray containing up to
five Carriers can be loaded into a Routine Bay. The number of Routine Bays varies depending on the
system configuration;
•
ARCHITECT c8000
®
System
configuration
:
six Routine Bays
•
ARCHITECT i2000
SR
®
System
configuration: four Routine Bays
•
ARCHITECT ci8200
®
System
configuration: twelve Routine Bays.
Component
Description and Function
Routine Bay and Bay Door
• One replaceable unit consists of two bays
• Number of bays depends on system configuration
• Load one Carrier Tray per bay
- Carrier Tray can hold up to five Carriers
• Slots that interface with the Carrier Section Alignment Sensor
• Locks in place to prevent user access to interior with anything other than a
Carrier Tray
Carrier/Tray Sensor
• Replaceable
• Detects the presence of a Carrier Tray
Standard Indicator and Unit
Detect Board (UDB)
• Provide status information on bay and Carrier Tray samples
• Contains rotary switch potentiometer set for corresponding bay (2) set
• 5VDC from SIB (fuse F5)
• LED is replaceable
• Connects to:
- UDB/PLSB
- UDB
- LED
- Carrier/Tray Sensor
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...