MODULE 4: POWER AND CIRCUIT BOARDS
4-12
ARCHITECT
®
RSH Service Training Guide
FOR INTERNAL USE ONLY
201532-102
Miscellaneous RSH Power Distribution
Power is distributed to three additional main RSH areas for use. They are:
•
Carrier Positioner Assembly
•
Carrier Transport Assembly
•
Sensor Interface Board (SIB)
Carrier Positioner Assembly
The Carrier Positioner Assembly receives and delivers:
•
+5VDC to the Encoder Sensor
Each PM has one positoner. On the
ci
8200
®
System is important to properly connect the cables to the
correct PM to ensure proper positioner function.
NOTE: The Carrier Positioner Board is replaceable. For additional information refer
to the ARCHITECT RSH Service and Support manual.
Carrier Transport Assembly
The Carrier Transport Assembly receives:
•
+5VDC for two boards and ten sensors
-
Carrier Transport Flex Board:
-
NOT replaceable
-
contains many sensors for positioning and recognition
-
Rail Guide Sensor Board:
-
replaceable
-
identifies when transport arm contains a Carrier