MODULE 4: POWER AND CIRCUIT BOARDS
ARCHITECT
®
RSH Service Training Guide
4-5
201532-102
FOR INTERNAL USE ONLY
POWER DISTRIBUTION
Power distribution to the RSH is dependent on system configuration.
ARCHITECT
®
c8000
®
System
On the
c
8000 System the Stepper Motor Driver/Sample Handler (SMD/SH) Power Supply provides power
to BOTH the
c
8000 System and the Retest Sample Handler (RSH) Card Cage. The RSH Card Cage is a
separate card cage from the main
c
8000 System. It is attached to the
c
8000 System right side during
system installation. The RSH Card Cage receives and distributes the following voltages:
Received Voltage
Supplied Voltage
Fuses
+5VDC
• Sensor Interface Board (SIB)
• Sample Handler Controller/CPU Board
• Indexer
• Motor Driver Board
• F1 (2A) in middle card cage area (
c
8000
System only)
+12VDC
• Card Cage Fans (
c
8000 System only)
• F2 (10A) in upper card cage area (
c
8000
System only)
+12VDC
• Sensor Interface Board
• Sample Handler Controller/CPU Board
• Indexer
• Motor Driver Board
N/A
+36VDC
• Motor Driver Board:
- Carrier Transport X axis motor
- Carrier Transport Theta motor
- Carrier Transport Z motor
N/A
J4 (+12VDC)
J1 (+36VDC)
J6 (+5VDC)
+3
6VD
C
c8000
®
SMD/SH Power Supply
c8000 System add on
Card Cage
+5
VD
C
+1
2VD
C
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...