MODULE 4 - POWER AND CIRCUIT BOARD ACTIVITY
Power and Circuit Board Procedures
Measure voltages on the Sensor Interface Board
(2.6 Block/Functional Diagrams) and
Record
recovered
voltages in the table below.
Card Cage Cabling
a.
List the components that receive 5VDC from the SIB.
Test Point
Expected
MEASURED
Comments
TP1
5VDC
4-5VDC
TP2
GND
-
TP3
5VDC
4-5VDC
TP4
5VDC
4-5VDC
TP5
GND
N/A
Fuse
Expected
MEASURED
Rating (Amp)
F1
5VDC
4-5VDC
1.5A
F2
5VDC
4-5VDC
1.5A
F3
5VDC
4-5VDC
4.0A
F4
5VDC
4-5VDC
4.0A
F5
5VDC
4-5VDC
4.0A
F6
3.3VDC
3-3.3VDC
1.0A
F7
5VDC
4-5VDC
1.5A
F8
3.3VDC
3-3.3VDC
1.0A
F9
5VDC
4-5VDC
1.5A
F10
3.3VDC
3-3.3VDC
1.0A
RSH Bar Code Reader
RSH Keypad
PLSB and UDBs
Carrier Positioner Board
Carrier Transport Sensors
Low Voltage Regulator
Input DC/DC
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...