MODULE 3: ROBOTICS
ARCHITECT
®
RSH System Service Training Guide
3-27
201532-102
FOR INTERNAL USE ONLY
REVIEW QUESTIONS AND CASE STUDIES
SCENARIO 1 During an
ci
8200
®
System installation the following error occurred during STARTUP
.
The following troubleshooting STEPs were performed and the issues was unresolved:
• Reseated and checked cables at SIB and transport
• Swapped Indexer Boards
What checks, measurements, procedures, etc. would you use to help isolate the root cause of the
failure? What would you do to try and resolve this error? What is the most likely cause of the
failure?
SCENARIO 2 Intermittent Carrier Rail Guide Sensor errors.
The following troubleshooting STEPs were performed, but the errors remain unresolved:
• Rail Guide Sensor Board was replaced
• Cables on SIB and transport were reseated
• Indexer Board was replaced
What checks, measurements, procedures, etc. would you use to help isolate the root cause of the
failure? What would you do to try and resolve this error? What do you think may be causing the
failure?
OK
?
Module: 009:14:2004 13:07:12
5409 Step loss detected on
(Carrier Transport X), actual
(expected), (0 (72)).
!
OK
?
Module: 009:14:2004 13:07:12
5409 Step loss detected on
(Carrier Transport X), actual
(expected), (0 (72)).
!
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...