MODULE 4: POWER AND CIRCUIT BOARDS
4-10
ARCHITECT
®
RSH Service Training Guide
FOR INTERNAL USE ONLY
201532-102
The
i
2000
SR
Card Cage receives and distributes the following voltages for RSH functions:
Received Voltage
Supplied Voltage
Fuses
+5VDC
• Sensor Interface Board (SIB)
• Sample Handler Controller/CPU Board
• Indexer
• Motor Driver Board
• F9 (7.5A) - CPU
• F16 (10A) - SIB
• F13 (2A) - 5VDC
+12VDC
• Sensor Interface Board
• Sample Handler Controller/CPU Board
• Indexer
• Motor Driver Board
N/A
+36VDC
• Motor Driver Board:
- Carrier Transport X axis motor
- Carrier Transport Theta motor
- Carrier Transport Z motor
• F14 (7.5A) - 36VDC
• F15 (2A) - 36VDC