![background image](http://html1.mh-extra.com/html/abbott/architect-rhs/architect-rhs_training-manual_450972057.webp)
MODULE 2: BASIC OPERATION
ARCHITECT
®
RSH Service Training Guide
2-17
201532-102
FOR INTERNAL USE ONLY
Priority Bay
The left most section of load/unload points on the RSH is referred to as the Priority Bay. Each location in
the Priority Bay that a Carrier can be loaded into is referred to as a section.
All system configurations have
one Priority Bay.
The Priority Bay has seven sections for loading carriers.
Component
Description and Function
Section and Section partition
• Seven sections (replaceable)
• Load one Carrier per section
• Slots that interface with the Carrier Section Alignment Sensor
Carrier/Tray Sensor
• Replaceable
• Detects the presence of a Carrier
Priority Indicator Lights and Sensor Board
(PLSB)
• Provide status information on section and sample
• One board for all seven sections
• LEDs are replaceable
• 5VDC from SIB (fuse F9)
• Connects to UDB, LED, Carrier/Tray Sensors
Connector between PLSB and UDB
Priority Bay Section
Carrier
Priority Status indicator light
Carrier/Tray Sensor
RSH Bar Code Reader
PLSB Board
Priority Bay
Summary of Contents for ARCHITECT RHS
Page 156: ...APPENDIX A 2 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 158: ...APPENDIX A 4 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 160: ...APPENDIX A 6 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...
Page 162: ...APPENDIX A 8 ARCHITECT RSH Service Training Guide FOR INTERNAL USE ONLY 201532 102 Notes Page ...