AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
45
Flowing reference gas
Gas inlet conditions as for sample gas
Test gases
Zero-point calibration
Sample-component-free process gas or substitute gas
End-point calibration
Process gas with a known sample gas concentration or substitute gas
Dew point
The dew point of the test gases must be approximately the same as the dew
point of the sample gas.
NOTE
The notes for calibrating (see page 233) must be considered.
Gas connections
see section "Caldos25: Gas connections" (see page 80)