AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
33
Version for use in Class I Div. 2 explosion hazard areas
Intended application
The AO2000 Series gas analyzers with analyzer modules Caldos25, Caldos27,
Limas21 UV, Limas21 HW, Magnos206, Magnos28, Magnos27 and Uras26 are
certified for use in Class 1, Div. 2 explosion hazard areas, Groups A, B, C, and
D, Temperature code T4.
Housing versions not equipped with conduit entries must be installed in a
suitable enclosure, cabinet or rack with provision for connection of Div. 2
wiring methods, acceptable to the local inspection authority having jurisdic-
tion.
Certificate
Certificate No. 1105720
Warnings
WARNING – EXPLOSION HAZARD
Substitution of components may impair suitability for Class I Division 2.
Secure all connectors with fasteners provided.
Relay contacts rated 30 V/1 A.
I/O connectors must only be connected to Class 2 circuits.
Ratings and ambient temperature
Analyzer module
Rating
Ambient temperature
Caldos25
max. 25 W
+5 to +45 °C
Caldos27
max. 17 W
+5 to +50 °C
Limas21 UV
max. 100 W
+5 to +45 °C
Limas21 HW
max. 100 W
+15 to +35 °C
Magnos206
max. 50 W
+5 to +50 °C
Magnos28
max. 50 W
+5 to +50 °C
Magnos27
max. 35 W
+5 to +45 °C
Uras26
max. 95 W
+5 to +45 °C