CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Waiting period following automatic calibration
If the
Output Current Response
parameter is set to Hold, current output is
halted for a specific time to allow the measurement value to stabilize after
automatic calibration is ended.
This interval is:
Test gas
→
Sample gas purge time + 4 x T90 or
Test gas
→
Sample gas purge time + 1 x T90-1 + 3 x T90-2.
Calibration data
Setting the calibration data is described in section "Calibration data for
automatic calibration" (see page 227).
Setting the T90 time constants is described in section "Filter parameteriza-
tion" (see page 163).