AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
149
Password protection
Elements of password protection
Password protection consists of three elements
•
Password level,
•
User group,
•
Password.
Password level
Each menu item is assigned a password level. Password levels are numbered
with 0, 1, 2 and 3.
Menu items are assigned to different password levels in order to assure that
specific menu items can only be changed by authorized users.
User group
The definition of a user group is that every user that belongs to it has access
entitlement at certain password levels, i.e. can make changes to the menu
items at these levels.
Some user groups are set-up at the factory.
A user group can be made up of one or more users.
Password
Every user group set-up in the system has a password.
The password consists of six digits which can be entered via the numeric
keypad.
Passwords are pre-assigned for the factory-set user groups.
Factory setting
User group
Access to password levels
Default password
Every user
0
None
Maintenance team
0, 1
471100
Specialist team
0, 1, 2
081500
Function block specialist 0, 1, 2, 3
325465
Field service
0, 1, 2, 3, 4, S
737842
It is recommended to change the default passwords.
CAUTION
After entering the password for password level 3, you can access all of the
function block applications. When configuring function blocks, existing
applications with their configurations and links can be damaged or de-
stroyed!
NOTE
Technical Bulletin "Function Blocks – Descriptions and configuration" con-
tains complete information on the "Function block" concept as well as de-
tailed descriptions of the individual function blocks.