270
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Changing analog output current range
Method
The current range of the individual analog outputs can be modified by set-
ting the parameter for the applicable
Analog output
function block
The
Technical Information "Function Blocks – Descriptions and configuration"
contains complete information on the individual function blocks..
Menu path
MENU
→
Configure
→
Function blocks
→
Outputs
→
Analog output
Changing current range
The current range is changed using the
Output current range
parameter.
Selection
The current range choices are 0 to 20 mA, 2 to 20 mA and 4 to 20 mA.
NOTE
The output signal may not be less than 0 mA and larger than 22 mA.
Current range limitation
The output signal is limited to the range which is specified in the parameters
Lower limit
and
Upper limit
. These parameters are set to 0 mA and
22 mA respectively in the factory.