AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
341
Pneumatic module: Troubleshooting
Flow error
Condensation in the flowmeter
•
Disconnect gas analyzer from the gas conditioning system.
•
Dry the flow meter by heating it and blowing it with compressed air.
•
Check the operation of the upstream sample gas cooler.
Inadequate gas supply
•
Connect the flow meter, ball chamber or pressure gauge directly to the
gas supply pump and measure the pressure or vacuum.
•
Check the pump and change the membrane if necessary.
•
Check the disposable filter and replace (see page 314) if necessary.
•
Check the solenoid valve(s) and replace if necessary.