AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
25
Technical data
Electrical specifications
Power supply
System housing with central unit and analyzer module: 100 to 240 V AC
(
−
15 %, + 10 %), 2,2 to 0,7 A, 47 to 63 Hz;
System housing with 2 analyzer modules: 24 V DC, max. 95 W per module,
Functional extra-low voltage "PELV"
Signal inputs and outputs
Functional extra-low voltage "PELV"
System bus, computer in-
terfaces
Functional extra-low voltage "PELV"
For further electrical data, see "Advance Optima AO2000 Series" data sheet.
System housing
Housing protection type
IP54 as per EN 60529
Ambient temperature
during operation with analyzer module installed in a system housing
without electronics module with electronics module
Limas11 IR
+5 to +45 °C
+5 to +45 °C, +5 to +40 °C with I/O modules
Uras26
+5 to +45 °C
+5 to +40 °C
Caldos25
+5 to +45 °C
+5 to +45 °C
Caldos27
+5 to +50 °C
+5 to +45 °C
Magnos206
+5 to +50 °C
+5 to +45 °C
Sample gas path ("containment system")
Sample gas
Flammable and non-flammable gases and vapors under atmospheric condi-
tions (p
abs
≤
1.1 bar, oxygen content
≤
21 vol.%); Temperature Class T4, for
Limas11 IR: T6; not potentially explosive under normal conditions, if poten-
tially explosive in the event of a fault, then only seldom and briefly (in ac-
cordance with Zone 2); (only for Limas11 IR:) occasionally potentially explosive
under normal conditions (in accordance with Zone 1), Temperature Class T4.
Mixtures of flammable gases and vapors and oxygen that are not explosive
under normal conditions or in the event of a fault. As a rule, these can be
achieved if the oxygen content is reliably limited to max. 2 vol.% or the
flammable component is reliably limited to max. 50 % LEL.
Flammable gases and vapors that are explosive under the conditions en-
countered during analysis, even when oxygen is excluded, may be present in
the mixture to be analyzed only in concentrations that are considered
non-critical according to safety regulations.
Flow
max. 40 l/h
Pressure
At sample gas inlet: max. positive pressure p
e
≤
3 hPa,
at sample gas outlet: atmospheric
Turning off the sample gas by user when shutting down the analyzer and in case of an alarm (pressur-
ized enclosure failure) additional special conditions for operation with
flammable sample gas (see page 27)