AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
189
Adding a system module
NOTES
If no system module is configured or an added system module is not yet
configured in a gas analyzer, the
NEW
softkey will appear on the screen. Press
this softkey to go directly to the
Setup System Modules
menu.
The procedure for adding an analyzer module and an external I/O device
differs from the procedure for adding an I/O module (see the following in-
structions).
Adding a new analyzer module or a new I/O device
1
Select the
Set Up System Modules
menu item.
The list of the system modules existing in the system is displayed.
2
Press the
NEW
softkey.
3
Enter the 14-digit serial number for the new system module.
4
In the list the added system module will be shown with
New
status.
5
Save the configuration change by pressing
ENTER
or discard it by
pressing
Back
.
Adding a new I/O module
1
Select the
Set Up System Modules
menu item.
The list of the system modules existing in the system is displayed.
2
Select the I/O module which has been added and automatically recog-
nized by the gas analyzer and press the
NEW
softkey.
3
When adding a digital I/O module:
Press the
FB APPL.
softkey and select a function block application.
4
In the list the new system module will be shown with
New
status.
5
Save the configuration change by pressing
ENTER
or discard it by
pressing
Back.
NOTE
When retrofitting a Profibus module, it must always be installed as the low-
est I/O module, i.e. on slot X20/X21.