CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Pressure regulator fault
Unstable pressure values
•
Adjust external pressure of the supply gases such that the output varia-
ble
for instrument air (
output
) is approx. 60 %,
for combustion air (
Air
) is approx. 55 %,
for combustion gas (
H2
) is approx. 42 %.
•
Have the pressure regulator modules checked.
Pressure regulator output variables not equal to set points
Air
Output variable
≤
50 %
•
Reduce combustion air primary
pressure.
Output variable
≥
90 %
•
Increase combustion air primary
pressure
H2
Output variable
≤
40 %
•
Reduce combustion gas primary
pressure.
Output variable
≥
90 %
•
Increase combustion gas primary
pressure.
Input
Output variable
≤
50 %
•
Reduce sample gas inlet pressure
Output
Output variable
≤
50 %
•
Increase the instrument air pressure
•
Clean air jet injector (see page 281).
•
Reduce length of the gas discharge
line or increase cross-section.
Output variable
≥
90 %
•
Reduce instrument air pressure.
Zero drift
Sample gas line contaminated
•
Clean sample gas line.
Combustion air catalyst is not operating adequately
•
Reduce hydrocarbon content.
•
Exchange catalyst.
Combustion gas line contaminated
•
Clean combustion gas line.