AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Limas21 HW: Operating specifications
Linearity deviation
Exhaust gas and diluted exhaust gas measurement:
≤
1 % of span,
≤
2 % of
measured value acc. to EPA specifications for automotive exhaust gas
measurement
Process measurement:
≤
1 % of span
Repeatability
≤
0.25 % of span
Zero drift
Exhaust gas and process measurement:
≤
1 ppm or
≤
1 % of span per
24 hours
Diluted exhaust gas measurement:
≤
250 ppb or
≤
2 % of span per 8 hours
in each case based on the smallest recommended measurement range
Daily automatic zero-point calibration recommended
Sensitivity drift
≤
1 % of the measured value per week
Output fluctuation (2
σ
)
Exhaust gas measurement:
≤
400 ppb or
≤
0.4 % of span at electronic T90
time = 5 s
Diluted exhaust gas measurement: NO
≤
50 ppb or
≤
0.5 % of span,
NO
2
≤
60 ppb or
≤
0.5 % of span at electronic T90 time = 15 s
Process measurement:
≤
150 ppb or
≤
0.15 % of span at electronic T90 time =
30 s
Detection limit (4
σ
)
Exhaust gas measurement:
≤
800 ppb or
≤
0.8 % of span at electronic T90
time = 5 s
Diluted exhaust gas measurement: NO
≤
100 ppb or
≤
1 % of span,
NO
2
≤
120 ppb or
≤
1 % of span at electronic T90 time = 15 s
Process measurement:
≤
300 ppb or
≤
0.3 % of span at electronic T90 time =
30 sec
Flow influence
Flow rate in the 20 to 90 l/h range: within the detection limit
Associated gas effect /
cross sensitivity
The knowledge of the sample gas composition is necessary for the analyzer
configuration. Selectivity measures to reduce associated gas effect (option-
al): Internal electronic cross-sensitivity correction or carrier gas correction
for a sample component by other sample components measured with the
Limas21 HW
Temperature influence
Ambient temperature in permissible range, sample cell thermostat control to
+82 °C;
at zero point:
≤
2 % of span per 10 °C,
on sensitivity:
≤
2 % of measured value per 10 °C
Air pressure effect
At zero point: no influence effect
On sensitivity with pressure correction with an integral pressure sen-
sor:
≤
0.2 % of the measured value per 1 % air pressure change
The connection of the pressure sensor is led to the outside. Pressure sensor
working range: p
abs
= 600 to 1250 hPa
Power supply influence
24 V DC ± 5 %:
≤
0.2 % of span
T
90
time
T90
≤
5 s for measurement cell length = 260 mm and sample gas flow =
60 l/h with non-linear filter (static/dynamic) = 15/1 s.
Low-pass time constant adjustable from 0 to 30 s.