CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
323
No. Text
S A W F Reaction/Remark
116 The Profibus Module is mount-
ed on the wrong slot! The Pro-
fibus interface is therefore not
working. Please remount the
Profibus Module on slot
X20/X21.
x x See message text
117 The configuration backup was
saved.
118 The configuration backup was
loaded. The system has been
restarted.
119 The system configuration could
not be loaded!, This system
therefore contains no configu-
ration now. Please enter menu:
Configure/System/Save
Save configuration to load your
backup configuration. Or use
SMT to re-install your configu-
ration.
x x See message text
System bus
201 The selected system bus mod-
ule could not be found.
x x Check plug connections and terminating resistors on the
system bus. Make sure the system bus module serial
number is correctly entered:
MENU
→
Diagnostics/Information
→
System
overview
203 The selected system bus mod-
ule does not exist.
x x Check plug connections and terminating resistors on the
system bus.
208 The system bus was not able to
transfer data into the data-
base.
x x The system bus module software version is not com-
patible with that of the system controller; update the
system controller software.
209 The system bus connection
to this module is interrupted.
x x Check the system bus connection to the indicated sys-
tem bus module. Check the power supply system of the
indicated system bus module.
210 The system bus module con-
figuration has changed.
x x For information; the configuration data are automati-
cally updated
211 The system bus module has no
more on-board memory.
x x Check the system bus module configuration:
MENU
→
Diagnostics/Information
→
System
overview
214 The system is currently main-
tained with Optima SMT.
215 The analyzer module has an
internal communication error!
x x Notify Service.
216 The analyzer module has an
internal program error!
x x Notify Service.
250 The selected analyzer module
could not be found!
x x Check the connectors and cabling.
251 The connection to the analyzer
has been lost!
x x Check the connectors and cabling.
252 The EEPROM data of the ana-
lyzer is faulty!
x x Check the configuration with TCT.
253 Communication with the ana-
lyzer is faulty!
x x Check the connectors and cabling.