AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
257
Calibration data for an analyzer module with 1 sample component
Example: Test gas = air
Calibration data for manual calibration
Calibration method
Common single-point calibration
Calibration measurement range
0 to 25 vol.% of O
2
Test gas concentration
20.96 vol.% of O
2
Calibration data for automatic calibration
Calibration method
Common calibration (test gas)
Single zero-point calibration
Always
Single span calibration
Never
Common zero and span calibration
Never
Test gas concentration zero gas
20.96 vol.% of O
2
Test gas concentration span gas
n.a.
Calibration data for an analyzer module with > 1 sample component
Example: Test gas = air
Calibration data for manual calibration
Calibration method
Substitute gas calibration
Zero component
O
2
in N
2
Zero range
0 to 25 vol.% of O
2
Span component
n.a.
Span range
n.a.
Test gas concentration zero gas
20.96 vol.% of O
2
Test gas concentration span gas
n.a.
Calibration data for automatic calibration
Calibration method
Substitute gas calibration
1)
Single zero-point calibration
Always
Single span calibration
Never
Common zero and span calibration
Never
Test gas concentration zero gas
20.96 vol.% of O
2
Test gas concentration span gas
n.a.
1)
Settings of sample components and measurement ranges for zero and
span calibration are taken over from the calibration data for manual
calibration.