AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Changing the number of decimal places
Menu path
MENU
→
Configure
→
Component-Specific
→
Measurement Range
→
Select Component
→
...
Selection
All measurement ranges configured (at the factory) for a sample component
are displayed.
Procedure
Select the measurement range with the arrow keys, press
SET DECIMAL
, set
the number of decimal places with the arrow keys and press
ENTER
to con-
firm.
NOTES
The setting only affects the values indicated on the screen.
The number of decimal places cannot be increased.
The altered number of decimal places is shown on the screen after switching
to measurement mode.
Number of decimal places
When the screen displays the measured value in physical units (e.g. ppm) the
number of places after the decimal point depends on the size of the set
measurement range:
Span
Places after the decimal point
≤
0.05
5
≤
0.5
4
≤
5
3
≤
50
2
≤
500
1
> 500
0
For the display of the measured value as a percentage of the whole measur-
ing range (%Span) two places are always shown after the decimal point.
When setting the parameters, the number of decimal places is the same as
in the display in measuring mode.