AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
311
Functions of the softkeys
New Lamp
Optimizes all receiver signals for all sample components;
this overwrites any stored initial intensities with a new
initial value.
Optimization
Optimizes receiver signals for a specific sample compo-
nent; this does not overwrite the stored initial intensi-
ties.
Optimize All
Optimizes receiver signals for all sample components;
this does not overwrite the stored initial intensities.
Perform amplification optimization
1
Turn on the zero gas supply.
If a solenoid valve is used to switch to zero gas, the supply will be acti-
vated automatically.
2
Select the
Amplification optimization
menu item.
3
Select the first sample component for which status message No. 301 is
present.
4
Press the
New Lamp
or
Optimize
or
Optimize All
softkey.
After pressing
New Lamp
a window will appear for entering the serial
number of the new lamp. After this the amplification optimization pro-
cedure will run automatically for all sample components and cannot be
interrupted.
5
Wait until the measurement value reading stabilizes and activate the
alignment procedure with
ENTER
.
6
Confirm the amplification optimization by pressing
ENTER
(the zero
point will be automatically aligned) or reject the result by pressing
Back
or
Meas
.
7
Repeat steps 3 to 6 for all sample components for which status mes-
sage No. 301 is present.