Calibrating the gas analyzer
Gas analyzer manual calibration
NOTES
Calibration should only be started after the warm-up phase.
A manual zero calibration must always be carried out before a manual span
calibration.
Gas analyzer manual calibration
1
Select the
Manual Calibration
menu:
MENU
→
Calibrate
→
Manual Calibration
2
For single calibration: Select
Component
and
Measurement Range
.
Zero calibration:
3
Select
Zero Gas
.
4
Turn on the zero gas supply.
5
If necessary, change the test gas concentration
1
shown,
ENTER
.
6
When the sample value indication stabilizes, initiate zero calibration
with
ENTER
.
7
Accept the calibration result with
ENTER
or
REPEAT
2
calibration (back to step 5)
or reject calibration with
Back
(back to step 6)
or reject calibration with
Meas
(back to measurement value readout).
Span calibration:
8
Select
Span Gas
.
9
Turn on the span gas supply.
10
If necessary, change the test gas concentration
3
shown,
ENTER
.
11
When the sample value indication stabilizes, initiate span calibration
with
ENTER
.
12
Accept the calibration result with
ENTER
or
REPEAT
4
calibration (back to step 10)
or reject calibration with
Back
(back to step 11)
or reject calibration with
Meas
(back to measurement value readout).
13
For single calibration, repeat steps 2 to 12 for other components and
measurement ranges.
1
The parameterized test gas concentration is displayed. If the set point is altered
here, the parameterized test gas concentration is overwritten.
2
A calibration may have to be repeated if the measured value is not stable after initi-
ation of the calibration. The repeated calibration is based on the measured value
obtained in the preceding calibration.
3
The parameterized test gas concentration is displayed. If the set point is altered
here, the parameterized test gas concentration is overwritten.
4
A calibration may have to be repeated if the measured value is not stable after initi-
ation of the calibration. The repeated calibration is based on the measured value
obtained in the preceding calibration.