330
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
No. Text
S A W F Reaction/Remark
519 Preamplifier overflow error:
Calibration could not be per-
formed because of preamplifier
override.
520 Initial zero calibration started. For information
521 Initial zero calibration ended.
For information
522 Initial zero calibration inter-
rupted.
For information
523 Initial zero calibration incom-
plete. System bus communica-
tion fault during calibration.
For information
524 Initial zero calibration started.
x For information; not logged
525 Linearization impossible: Line-
arization did not produce a
valid result. Measurement
See message text
526 Linearization impossible: Line-
arization could not be per-
formed, i.e. the characteristic is
linear.
See message text
527 Initial calibration for compo-
nent:
For information
528 Autocalibration not started, i.e.
manual calibration was run-
ning.
For information
529 Calibration was stopped be-
cause no raw measured values
were recorded.
x x
530 Calibration
stopped
because
the pressure switch did not
detect any calibration gas.
x x
531 Automatic validation started.
For information
532 Automatic validation ended.
For information
533 Automatic validation externally
interrupted.
For information
534 Automatic validation in pro-
gress.
x For information; not logged
535 Automatic validation successful
for:
536 Automatic validation out of
limits for:
537 Automatic validation out of
limits for:
x
User-Configured Messages
800 An external error occurred dur-
ing:
x x Default texts for the
Message Generator
function block
are supplemented with the full text defined during func-
tion block configuration.
801 A user-defined error occurred
during:
x x
802 A user-defined maintenance
request occurred during
x
803 A user-defined maintenance
mode event occurred during:
x