Special requirements for operation
•
The gas analyzer must be switched off if it is obviously faulty (is not
operating smoothly).
•
The connectors may not be plugged in or unplugged while the power is
on.
•
The gas analyzer housing may not be opened while the power is on.
•
In order to comply with degree of case protection IP54,
•
Cables have to be properly inserted in the screwed cable glands and
sealed by screwing the nut firmly,
•
Cable connectors not in use have to be closed with suitable vent
plugs,
•
System bus and 24 VDC power supply terminals not in use have to be
closed with the provided seals.
•
Only the original battery types Varta CR2032 type no. 6032 or Renata
type no. CR2032 MFR may be used as replacement for the battery on the
system controller.