Z E S ZIMMER Electronic Systems GmbH Chapter 5
99
c2
same as c1
c1&c2 both cursors are moved at the same time.
Split
after pressing this button the display will be split in four separated graphs, refer the
following picture:
Figure 29: The plot display; split on
Hint: the settings of the cursor are the same for all four to get accurate readings at the selected
moment for all channels.
Fit
By this the actual selected graph (A to D) is rescaled to fit into the screen.
5.4.6.3 Vec.
A/B
The settings in group A and B are identical, so they are just described once.
On this tabs you see the vectors of voltage and current (Fresnel diagramm). In the lower left
corner you see the amplitude and the phase angle of the selected signal. In the upper right
corner you see the rotating direction of the vectors and the phase counting of the voltages.
‘1,2,3’ means that the zero crossings are in the order 1
st
, 2
nd
, 3
rd
phase. ‘3,2,1’ means that you
have the opposite order. By pressing the rotary knob you can do the following settings:
Signl
Here you can choose the signal to be displayed. The possible values are voltage or
current.
Ref.
Here you can choose the reference for the angles:
U
U1 is set to 0°
I
I1 is set to 0°
None
The sync. signal is used as 0° reference.
All angles which are displayed in this menu are relative to this reference angle.
Zoom
You can zoom the selected signal to get longer vectors.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......