Z E S ZIMMER Electronic Systems GmbH Chapter 4
57
Pressing the rotary knob returns you to the tab selection.
4.4.2.2.3 Digital
Inputs
tab
In this menu you get the actual state of the digital inputs. The inputs 1 to 4 of each card are
only used for state indication. The inputs 5 and 6 can be used as state indicators or for
frequency and direction input (determined by the phase shift between input 5 and 6). In the last
case the input 5 is used to measure the frequency. This is multiplied with the ‘Scale’ value and
displayed under ‘Frequency’. A negative frequency value indicates a reverse rotation direction.
To change the scaling press the rotary knob and Scl.A (or Scl.B for the card B).
Figure 20: Digital inputs
Pressing the rotary knob returns you to the tab selection.
4.4.2.2.4 Digital
Outputs
tab
If you are in the setting mode of this menu, you can with the rotary knob select a channel.
Value
This allows you to set-up the value which should be output. See chapter 4.5,
‘Entering identifiers’ for details.
Cond
Here you set-up, under which condition the output is in the ‘alarm state’ (= high
impedance of output, symbolised lamp is on!):
on:
The output has always alarm state.
off:
The output has never alarm state.
>=:
The output has alarm state if the Value is bigger or equal to the Limit.
<:
The output has alarm state if the Value is smaller than the Limit.
Limit
Here you set-up, which limit is compared to the Value.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......