Z E S ZIMMER Electronic Systems GmbH Chapter 13
263
Output load: load resistance > 2k
Ω
13.9.3 Digital inputs
Input signal:
U
low
max=1V, U
high
min=4V@2mA, U
high
max=60V@3mA
13.9.4 Frequency
inputs
Input signal:
U
low
max=1V, U
high
min=4V, U
high
max=10V
Input resistance:
1M
Ω
Maximum frequency: 5MHz
Accuracy:
±
100ppm of measuring value
13.9.5 Digital outputs
Open collector outputs. See 14.6, ‘Functional block diagram processing signal interface’
Output high impedance:
max 30V@100
µ
A
Output low impedance:
max. 1.5V@100mA
13.9.6 Auxiliary
supply
Output voltage:
±
5V, 10% @ 50mA
13.9.7 Frequency/direction input
Incremental sensors without direction information
Connect them to FIn_1 only. Let FIn_2 open!
Incremental sensors with direction information
Connect the frequency track (usually track A) to FIn_1.
Connect the direction track (usually track B) to FIn_2.
A positive frequency is displayed, if the direction signal is high at the rising edge of the
frequency signal.
A negative frequency is displayed, if the direction signal is low at the rising edge of the
frequency signal.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......