Z E S ZIMMER Electronic Systems GmbH Chapter 10
139
5: 220V/60Hz
6: 240V/50Hz
7: 240V/60Hz
10.2.2.3.9 :VERSion
EDIT
SCPI:
:CALCulate:LIMit:VERSion <NRi>
SHORT: EDIT <NRi>
ID:
n/a
Mode:
prCE
Type:
long int
Suffix:
n/a
Value:
0...4
List:
n/a
Unit:
n/a
Group:
n/a
Sets or reads the edition of the harmonic standard:
0: EN61000-3-2:1995 and EN61000-4-7:1993 [*RST default value]
1: EN61000-3-2:1995 and EN61000-4-7:2008
2: EN61000-3-2:2006 and EN61000-4-7:1993
3: EN61000-3-2:2006 and EN61000-4-7:2008
4: EN61000-3-12:2005 and EN61000-4-7:2008
10.2.2.3.10 :ZREF
ZREF
SCPI:
:CALCulate:LIMit:ZREF <NRf>
SHORT: ZREF <NRf>
ID:
n/a
Mode:
Flicker
Type:
int
Suffix:
n/a
Value:
0, 1
List:
n/a
Unit:
n/a
Group:
n/a
Sets or reads the Z
ref
for the EN61000-3-11:2000 limit calculation.
0: (0.24+j0.15)
Ω
[*RST default value]
1: (0.40+j0.25)
Ω
10.2.2.3.11 :ZTESt
ZTST
SCPI:
:CALCulate:LIMit:ZTESt <NRf>
SHORT: ZTST <NRf>
ID:
n/a
Mode:
Flicker
Type:
float
Suffix:
n/a
Value:
0.05...1, [*RST default value] = 0.283
List:
n/a
Unit:
Ω
Group:
n/a
Sets or reads the Z
test
for the EN61000-3-11:2000 limit calculation.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......