Interfaces (IEEE option L45-O1) LMG450
224
Two comments on the syntax in the ‘Sent’ column: A ‘
∪
’ stands for a space character, a ‘
↵
‘
for the <cr> (carriage return) character (which is the enter key, if you use for example a RS232
terminal-program on your PC).
No.
Sent
Received
Comments
1
*rst
↵
Reset the instrument to it’s default values.
After this first command the instrument
changes to the remote state.
2
*idn?
↵
ZES ZIMMER Electronic
Systems GmbH, LMG450,
serial number, version
You ask the instrument for it’s identification.
3a
fetc:volt:trmsn?
↵
0
Reads out the voltage of channel n(=1,2,...).
You don’t get the actual value, because there
has been no values copied to the interface
buffer
3b
read:volt:trmsn?
↵
220.34
Waits until the end of the measuring cycle,
copies the values of channel n to the
interface buffer and outputs the actual
voltage
3c
fetc:volt:trmsn?
↵
220.34
No values were copied to the interface buffer,
so you get the same result!
4a
read:volt:trmsn?;:rea
d:curr:trmsn?
↵
220.21;0.6437
Waits until the end of the measuring cycle,
copies the values of channel n to the
interface buffer and outputs the actual
voltage. Then the instrument waits until the
end of the next measuring cycle, copies the
values to the interface buffer and outputs the
actual current. The voltage and current are
from different measuring cycles!!
4b
read:volt:trmsn?;:fet
c:curr:trmsn?
↵
221.13;0.6432
Waits until the end of the measuring cycle,
copies the values of channel n to the
interface buffer and outputs the actual
voltage and current. The voltage and current
are from the same measuring cycle!!
4c
read:volt:trmsn?
↵
fetc:curr:trmsn?
↵
217.75;0.6135
Waits until the end of the measuring cycle,
copies the values to the interface buffer and
outputs the actual voltage and current of
channel n. The voltage and current are also
from the same measuring cycle!!
5a
calc:form
∪
„Bpk=Ur
ect/(4*f*3*0.000091
6);
↵
Hpk=Ipp/2*3/0.0856
08;
↵
Ua=Bpk/1.2566e-
6/Hpk;“
↵
Enters the script between the „“ signs. This
script is stored in the script editor and
executed from the next measuring cycle.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......