Logging of values to drives, printer and interfaces LMG450
234
every cycle
The values are periodically output after each measuring cycle. Please make
sure, that the output device is fast enough!
periodic
The values are output after the interval you set up. The minimum time is 10s.
After you have chosen 'periodic' you get this time in a highlighted box. If the
value is ok, you press just Enter. If not press Per. and change it. Leave the
box with Enter.
every integral
The output interval depends on the settings of the Int. Time menu. The
outputs starts, if you have started an integration. For the different integration
modes you get:
continuous
Output at end of measuring cycle
interval
One time output after the integration time is over.
periodic
Periodic after each integration time
summing
Output at end of measuring cycle
by script
The output is done, when the
print()
function in the script editor is called (see
4.4.4.2.7, ‘Functions’).
11.4 Output
formats
Output as displayed
As default the values are output in the same position like at the display. A one-time output of
default menu, 1
st
channel could look like this:
Itrms:1= 0.0270 A
Utrms:1= 0.1414 V
P:1=-0.004 W
Q:1= 0.000 var
S:1= 0.004 VA
PF:1= 0.9992
Output as csv (Excel)
At periodic output it is an advantage, if the values are ordered by time. To do this choose the
'csv (Excel)' format.
dt/s,Itrms1/A,Utrms1/V,P1/W,Q1/var,S1/VA
216.00E-03, 289.94E-03, 221.61E+00, 41.313E+00, 49.210E+00, 64.253E+00
716.00E-03, 289.51E-03, 221.68E+00, 41.290E+00, 49.132E+00, 64.178E+00
1.2160E+00, 289.64E-03, 221.59E+00, 41.172E+00, 49.234E+00, 64.180E+00
1.7230E+00, 290.86E-03, 221.45E+00, 41.291E+00, 49.435E+00, 64.410E+00
2.2160E+00, 289.72E-03, 221.54E+00, 41.261E+00, 49.164E+00, 64.184E+00
2.7160E+00, 289.95E-03, 221.57E+00, 41.335E+00, 49.181E+00, 64.245E+00
3.2160E+00, 289.21E-03, 221.54E+00, 41.206E+00, 49.065E+00, 64.073E+00
3.7160E+00, 289.75E-03, 221.46E+00, 41.280E+00, 49.126E+00, 64.167E+00
4.2160E+00, 289.83E-03, 221.44E+00, 41.232E+00, 49.181E+00, 64.178E+00
4.7160E+00, 290.01E-03, 221.50E+00, 41.290E+00, 49.210E+00, 64.238E+00
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......