Sets or queries the setting of the signal coupling. Allowed values are:
‘0’ or ‘ACDC’ for AC+DC coupling [*RST default value]
‘1’ for AC coupling
10.2.8 :INSTrument
commands
Here general set-ups of the instrument are done.
:CALCulate
:DISPlay
:FETCh
:FORMat
:INITiate
:INPut
:INSTrument
→
:SELect
:MEMory
:READ
:SENSe
:SOURce
:STATus
:SYSTem
:TRIGger
10.2.8.1 :SELect
MODE
SCPI:
:INSTrument:SELect <NRi>
SHORT: MODE <NRi>
ID:
n/a
Mode:
All
Type:
long int
Suffix:
n/a
Value:
0...4
List:
n/a
Unit:
n/a
Group:
n/a
Sets or reads the measuring mode:
‘0’ or ‘NORML’ for normal measuring mode [*RST default value]
‘1’ or ‘CEHRM’ for CE harmonic measuring mode
‘2’ or ‘CEFLK’ for CE flicker measuring mode
‘3’ or ‘HRMHUN’ for 100 harmonics measuring mode
‘4’ or ‘TRANS’ for transient measuring mode
Hint
The execution of this command can take up to few seconds. The LMG works internally with a
watchdog protection. To prevent that the watchdog becomes active, the ‘MODE’ command
should be send as the only command in a message. Just the ‘*OPC?’ can be added to get a
feedback, if the command has finished (‘MODE x;*OPC?’).
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......