Z E S ZIMMER Electronic Systems GmbH Chapter 5
95
Start of integration
In general there are three ways to start an integration. You can either enter a start time with
Start Time, Start Date, or you can simply press the Start button or you can start via the
external sync jack of the LMG450 (see 14.1.1). The first cycle which is taken into account is
the cycle which follows the actual cycle.
The integration time should be an integer number of times the cycle time.
Stop of integration
The last cycle which is taken into account is the cycle when the Stop button or any other stop
signal appear.
State of integration
The integration can be in 6 different states:
Reset
The energy measurement is stopped, the values are reset to 0
Wait
If the start time is later than the actual time and you have started the integration this
state appears until the start time is reached.
Start
This state is displayed from the logical start of integration (e.g. pressing the Start
button) until the physical start of integration which is always the begin of the next
cycle.
Run
This is displayed while the physical integration is running
Stop
This state is displayed from the logical end of integration (e.g. pressing the Stop
button) until the physical end of integration which is always the end of the actual
cycle.
Hold
This is displayed if the integration has finished. The integrated values are hold, until
the integration continuous (only summing mode) or the values are reset by RESET
or Start.
The logical integration is running, if the state is displayed inverse.
5.4.6 Graphical
display
With
Graph you see the graphical display of the normal measuring mode. Also in this menu we
have the group concept. With the rotary knob you can select different tabs (‘Scope A’, ‘Scope
B’, ‘Vec.A’, ‘Vec.B’ and ‘Extended’).
5.4.6.1 Scope
A/B
The settings in group A and B are identical, so they are just described once.
Left of this graph you see the Y scaling (ydiv), the Y scaling factor (yzoom) and the four
selected signals. Under the graph you see the start position of the graph in seconds, the X
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......