CE-Flicker measuring mode (option L45-O4) LMG450
114
Per.
This is the number of short term periods for the long term measurement. The standard
value is 12 periods for a long term time of 2 hours.
Figure 35: Measuring menu in CE-Flicker mode
7.1.2 Ztest/Zref
tab
This is only available when EN61000-3-11 is selected.
With Ztest you have to enter your actual used test impedance in
Ω
.
With Zref you have to choose, if your reference impedance is (0.24+j0.15)
Ω
or (0.4+j0.25)
Ω
.
The values for d
c
, d
max
, P
st
and P
lt
are recalculated as described in EN61000-3-11:2000
7.2 Measuring
ranges
(Range)
The settings are the same like in ‘5.2, Measuring ranges (Range)’. Please note that the
standards require a continuous measuring without gaps. For that reason the auto range
function should be deactivated to prevent a range change, because this will cause a short
measurement with invalid values.
7.3
Definition of measuring values
The following basic definitions are used:
n
The harmonic order.
T
The time of an integer number of the periods of the synchronisation signal. The integer
factor depends on the standard IEC61000-4-7. Actually 16 periods are measured.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......