Z E S ZIMMER Electronic Systems GmbH Chapter 10
163
Reads the actual result of the current check:
Bit 0: Set if the total current evaluation failed (=if any of the sub evaluation failed). Cleared
otherwise.
Bit 1: Set if any of the harmonics was > 100% of the allowed limit, cleared otherwise.
Bit 2: Set if the fluctuating harmonics were for more than 10% of the 2.5 minute window
between 100% and 150% of the limit. Cleared otherwise.
Bit 3: Set if questionable vales (with ‘?’) have occurred. Cleared otherwise.
Bit 10: Set if any harmonic is > 150% of limits. Cleared otherwise.
10.2.4.1.7.6.17 :THARmonic?
HTHC?
Ithc
SCPI:
:FETCh[:SCALar]:HARMonics:CURRent:THARmonic? /qonly/ |
:READ[:SCALar]:HARMonics:CURRent:THARmonic? /qonly/
SHORT: HTHC? /qonly/
ID:
Ithc
Mode:
prCE
Type:
float
Suffix:
1...4
Value:
n/a
List:
n/a
Unit:
A
Group:
n/a
Reads the Total Harmonic Current.
10.2.4.1.7.6.18 :THDistort?
HIHD?
Ithd
SCPI:
:FETCh[:SCALar]:HARMonics:CURRent:THDistort? /qonly/ |
:READ[:SCALar]:HARMonics:CURRent:THDistort? /qonly/
SHORT: HIHD? /qonly/
ID:
Ithd
Mode:
prCE, Flicker, HARM100
Type:
float
Suffix:
1...4/12
Value:
in %
List:
n/a
Unit:
n/a
Group:
n/a
Reads the THD of the current.
10.2.4.1.7.7 :LTRemain?
HLTR?
SCPI:
:FETCh[:SCALar]:HARMonics:LTRemain? /qonly/ |
:READ[:SCALar]: HARMonics:LTRemain? /qonly/
SHORT: HLTR? /qonly/
ID:
n/a
Mode:
prCE
Type:
long int
Suffix:
n/a
Value:
n/a
List:
n/a
Unit:
s
Group:
n/a
Reads the remaining long time for the harmonic measurement.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......