CE-Flicker measuring mode (option L45-O4) LMG450
116
7.4.2 Voltage
Here you see the TRMS value, the THD and the frequency.
7.4.3 Current
Here you see the TRMS value, the THD and the frequency.
7.4.4 Power
For the display of the values we have the same rules like in the normal measuring mode (see
5.4, ‘Display of values’)
7.4.5
Flicker (Int. Val)
In this menu you see the flicker values of the equipment under test (EUT). You see the short
term flicker level P
st
, the long term flicker level P
lt
, the actual flicker level P
mom
, the relative
steady-state voltage change d
c
, the maximum relative voltage change d
max
, the remaining long
term time, the remaining short term time and the state of the flicker measuring.
The state can be ‘starting’ (8s from pressing START), ‘running’ (while the long term interval)
and ‘stooped’ after the measuring.
Figure 36: Evaluation of flicker measurement
d
max
is measured over the long term time.
d
c
is the relative voltage change between two ‘constant’ voltages. Therefore it can have
different values:
d
c
=------- means that there was no constant voltage.
d
c
=0.000% means that there was exactly one constant voltage.
Any other value is the biggest difference between two constant voltages.
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......