Interfaces (IEEE option L45-O1) LMG450
152
ID:
dcs
Mode:
Flicker
Type:
float
Suffix:
1...4
Value:
in %
List:
n/a
Unit:
n/a
Group:
n/a
Reads the d
c
value of the source.
10.2.4.1.5.4.3 :DELTat?
FSDT?
dts
SCPI:
:FETCh[:SCALar]:FLICker:SOURce:DELTat? /qonly/ <list>|
:READ[:SCALar]:FLICker:SOURce:DELTat? /qonly/ <list>
SHORT: FSDT? /qonly/ <list>
ID:
dts
Mode:
Flicker
Type:
float
Suffix:
1...4
Value:
in %
List:
0...31
Unit:
n/a
Group:
n/a
Reads the d(t) values of the source. After each measuring cycle over 16 periods you can get 32
values.
10.2.4.1.5.4.4 :DMAX?
FSDX?
dmaxs
SCPI:
:FETCh[:SCALar]:FLICker:SOURce:DMAX? /qonly/ |
:READ[:SCALar]:FLICker:SOURce:DMAX? /qonly/
SHORT: FSDX? /qonly/
ID:
dmaxs
Mode:
Flicker
Type:
float
Suffix:
1...4
Value:
in %
List:
n/a
Unit:
n/a
Group:
n/a
Reads the d
max
value of the source.
10.2.4.1.5.4.5 :DTMViolation?
FSMV?
SCPI:
:FETCh[:SCALar]:FLICker:SOURce:DTMViolation? /qonly/ |
:READ[:SCALar]:FLICker:SOURce:DTMViolation? /qonly/
SHORT: FSMV? /qonly/
ID:
n/a
Mode:
Flicker
Type:
long int
Suffix:
1...4
Value:
n/a
List:
n/a
Unit:
n/a
Group:
n/a
Reads the maximum number of half waves for which the d(t) of the source was bigger than
allowed in the standard.
10.2.4.1.5.4.6 :HWTRms?
FSRM?
Uhws
SCPI:
:FETCh[:SCALar]:FLICker:SOURce:HWTRms? /qonly/ <list> |
Содержание LMG450
Страница 1: ...4 Channel LMG450 Power Meter User manual Status 2009 02 17...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 36: ......
Страница 76: ......
Страница 102: ......
Страница 112: ......
Страница 118: ......
Страница 124: ......
Страница 230: ......
Страница 240: ......
Страница 270: ...System design LMG450 270 14 2 Functional block diagram LMG450 Figure 47 Functional block diagram LMG450...
Страница 292: ......