8 Troubleshooting | 8.3 Chamber
ZEISS
à
INFO: If initialization of the stage does not solve the stage problem, contact your local
ZEISS service representative.
8.3.2 Defining the Post Initialization Position of the Stage
Purpose
You can configure the position to which the stage drives after the initialization procedure. Other-
wise, the stage drives to the center position.
Prerequisite
ü
Requires the Supervisor privilege.
1. From the Windows start menu, select SmartSEM > SmartSEM Administrator.
à
The SmartSEM Administrator Log on window is displayed.
2. Enter user name and password.
3. To confirm, click OK.
à
The SmartSEM Administrator window is displayed showing the user list.
4. Click Column/Stage.
5. In the Stage Post Initialisation Position input fields, enter the desired position.
Alternatively, use the dual joystick to navigate to the desired position and click Set to cur-
rent position.
6. To activate the function, activate the Post Init. Posn Valid checkbox.
8.3.3 Changing the Joystick TV Angle
Purpose
In TV mode (chamberscope), it can occur that dual joystick and stage seem to move to opposite
directions. This is because the selected CCD camera is installed at a certain angle relative to the
stage. Thus, the camera shows a side-inverted view. To remedy this, you need to change the joy-
stick TV angle setting in the software.
Info
Prerequisite
ü
Requires the Supervisor privilege.
1. From the Windows start menu, select SmartSEM > SmartSEM Administrator.
à
The SmartSEM Administrator Log on window is displayed.
2. Enter user name and password.
3. To confirm, click OK.
4. The SmartSEM Administrator window is displayed showing the user list.
5. Click Column/Stage.
6. In the Stage Options section, double-click the Joystick TV Angle input field.
7. Enter an angle depending on the installation lo-
cation of the CCD camera.
If the CCD camera is installed at the back, enter
180°
.
If the CCD camera is installed at the front, enter
0°
.
If the CCD camera is installed at the side, enter
90°
.
180°
0°
90°