10 Technical Data and Conformity | 10.2 Installation Requirements
Parameter
Requirement
Protective ground
High leakage currents are present in the microscope. Therefore, the
microscope has to be connected to a separate protective ground.
An exclusive grounding connection to earth must be provided as part
of the building installation, i.e. a grounding screw terminal (Ø 8 mm)
which is directly connected to the PE of the FEMALE RECEPTACLE as
short as possible.
This grounding connection must not be common to other electrical
equipment.
A grounding wire AWG10 (≥ 5 m) is delivered with the microscope. It
serves as connection between the grounding screw terminal (Ø 8 mm)
and the microscope (Ø 6 mm).
Cross section
: Min. AWG10 (between grounding screw terminal and
PE of the CEE FEMALE RECEPTACLE)
2
Cooling Water
Major components of the microscope such as electron-optic lenses, parts of the electronics and
the turbo molecular pump are water-cooled. Any cooling solution has to fulfil the following re-
quirements.
Parameter
Requirement
Water flow rate
60–70 l/h
Pressure
Adjustable up to 3 bar
Water tempera-
ture
20–22 °C
Stability
0.5 °C/10 min
Heat dissipation
1 kW
Connection hose
6 mm inside diameter. Two pieces 10 m each are delivered.
Parameter
Requirement
Water flow rate
> 1.5 l/min
Pressure
2–3 bar
Water tempera-
ture
18–22 °C
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000
119