ZEISS
2 Safety | 2.4 Safety Equipment
Position and Figure of the Safety La-
bel
Description
8
At the rear of the microscope
WARNING
High leakage current
347800-0033-000-07en
Ensure proper grounding.
Do not operate the electron
microscope without a separate
ground connection.
High leakage current
Ensure proper grounding.
Do not operate the electron microscope without
a separate ground connection.
Reorder no. 347800-0033-000-07en
9
At the rear of the microscope
Radiation hazard
Do not remove any parts.
Observe local safety and X-ray protection regula-
tions.
Service only through authorized personnel.
Reorder no. 347800-0033-000-06en
Liquid nitrogen
CAUTION
347800-0033-000-09en
Contact with liquid or cold vapors
can cause freeze burns or severe
frostbite.
Wear face, eye and skin protection.
For relevant options:
Liquid nitrogen
Contact with liquid or cold vapors can cause
freeze burns or severe frostbite.
Wear face, eye and skin protection.
Reorder no. 347800-0033-000-09en
Underneath the cover panels of the microscope, there are some more safety labels, which are ad-
dressed to authorized ZEISS service representatives. These safety labels are described in the docu-
ments for ZEISS service representatives.
Position and Figure of the Type Plate
Description
At the rear of the microscope
Type plate
Position and Figure of the Legal In-
formation
Description
At the rear of the microscope
Caution
X-rays are produced in this Scanning Electron Mi-
croscope!
The acceleration voltage is limited to 30 kV.
Observe local safety and X-ray protection regula-
tions.
Order no. 346000-0088-000
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000