17-13
IM 253710-01E
Specifications
17
Item
Specification
Accessories
Power cord : 1 piece
Spare power fuse : 1 piece (Attached to the fuse holder)
Cover plate: 4 pieces (Attach cover plates on unused slots using screws (see below).)
Current input protective cover: 1 piece.
Screw: 20 pieces (M3, length: 5mm, for fixing cover plates and current input protective covers in
place).
Printer roll paper : 1 roll, only provided with option /B5
Rubber feet : 2 sets
User’s manual : 1 piece, this manual
Communication Interface User’s Manual : 1 piece
Safety standard
*1
Complying standard
EN61010-1 (apply for 253710, 253751, 253752, 253771)
Overvoltage category (Installation category) II
*2
Measurement category II*
6
Pollution degree 2
*3
Emission
*1
Complying standard
• EN61326 Class A(apply for 253710, 253751, 253752, 253771)
• EN55011 Class A, Group1 (apply for 253710, 253751, 253752, 253771)
This product is a Class A (for industrial environment) product.
Operation of this product in a residential area may cause radio
interference in which case the user is required to correct the interference.
• EN61000-3-2 (apply for 253710, 253751, 253752, 253771)
• EN61000-3-3 (apply for 253710, 253751, 253752, 253771)
• C-tick EN55011 Class A, Group1 (apply for 253710, 253751, 253752, 253771)
Cable requirement
• Measuring Input
To bundle the wires between source and load for each phase and to
separate the input signal wires by less than 50 mm between each phase
and neutral line.
Attach a ferrite core (YOKOGAWA: A1179MN) to the cable near the
terminal. However, this is not necessary if the input module is 253771.
• External trigger input terminal
Use a 3D2W BNC cable.
*4
• Trigger output terminal
Same as the above external trigger input terminal.
• External clock input terminal
Same as the above external trigger input terminal.
• RS-232 connector
Use an RS-232 shielded cable.
*4
• Centronics interface connector
Use a shielded cable*
4
for the connection.
• SCSI connector
Use a SCSI shielded cable.
*4
• GPIB connector
Use a GPIB shielded cable
*4
and attach a ferrite core (YOKOGAWA:
A1179MN) on the cable near the terminal.
Immunity
*1
Complying standard
EN61326 Annex A
*5
(apply for 253710, 253751, 253752, 253771)
Cable requirement
Same as the above cable requirement about emission.
*1 Emission, immunity and safety standards apply to products having the CE Mark. For all other products, please contact your
nearest YOKOGAWA representative as listed on the back cover of this manual.
*2 Overvoltage Categories define transient overvoltage levels, including impulse withstand voltage levels. Overvoltage
Category I: Applies to equipment supplied with electricity from a circuit containing an overvoltage control device.
Overvoltage Category II: Applies to equipment supplied with electricity from fixed installations like a distribution board.
*3 Pollution Degree: Applies to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface
resistivity. Pollution Degree 1: Applies to closed atmospheres (with no, or only dry, non-conductive pollution). Pollution
Degree 2: Applies to normal indoor atmospheres (with only non-conductive pollution).
*4 The cable length is less than 3 m.
*5 Annex A (normative) : Immunity test requirements for equipment intended for use in industrial locations.
*6 Measurement category II (CAT II) applies to measurement of electrical equipment which is supplied from fixed installations
such as a wall outlet wired from a distribution board, or of the wires themselves.
17.13 General Specifications