14-9
IM 253710-01E
External Trigger Output and Other Operations
14
Turning ON/OFF the action used to save the numerical data, waveform data, and
setup parameters
3.
Press the [Save to File] soft key to select [ON] or [OFF].
Turning ON/OFF the action used to output (or save) the screen image data
4.
Press the [Hard Copy] soft key to select [ON] or [OFF].
Setting the number of times to perform the action
5.
Turn the jog shuttle to set the action count.
Executing/Stopping action-on-trigger
Press the START/STOP key to execute action-on-trigger. Action-on-trigger stops
when the START/STOP key is pressed again to stop the data acquisition or after
performing the action the specified number of times. When the SINGLE START key
is pressed, the action-on-trigger is carried out once.
14.4 Setting the Action-on-Trigger