14-10
IM 253710-01E
Explanation
You can have a specified action to be carried out when a trigger occurs.
Saving the numerical data, waveform data, and setup parameters
You can select whether or not to save the data specified in chapter 12 (numerical
data, waveform data, or setup parameters) when a trigger occurs. The save
destination is the medium specified in chapter 12.
• ON: Saves the data.
• OFF: Does not save the data.
Outputting or saving screen image data
You can select whether or not to output (or save) the screen image data when a
trigger occurs. The output (save) destination is the destination specified in chapter
13.
• ON: Outputs (or saves) the data.
• ON: Does not output (or save) the data.
Action count
• 1 to 65536
Action-on-trigger terminates after performing the action the specified number of
times.
• Infinite
The action is repeated until the data acquisition is stopped with the START/STOP
key.
• The action count cannot be changed while action-on-trigger is in progress.
Executing/Stopping action-on-trigger
• Executing action-on-trigger
Action-on-trigger is executed when waveform acquisition is started with the
START/STOP or SINGLE START key. When the SINGLE START key is pressed,
the action-on-trigger is carried out once.
• Stopping action-on-trigger
Press the START/STOP key to stop the data acquisition.
Name of the file for saving data
• Files are automatically named with three digit numbers from 000 to 999. You can
specify a common name (up to five characters) that is placed before the three digit
number. The common name (up to five characters) is the file name that is
specified during the save operation of various data (see chapters 12 and 13).
• If a file with the same name exists at the destination, an error message is displayed
and the file is not saved.
Note
The number of files that can be saved varies depending on the data size of the file and the
free space at the save destination.
14.4 Setting the Action-on-Trigger